Agilent 1260 Infinity II Preparative Autosampler User Manual 143
Error Information
8
Sampler Error Messages
Valve switch not found.
Error ID: 34319
The injection valve has failed to switch into the position. The valve position
sensor monitors the position of the valve. If the valve fails to switch into the
position (mainpass or bypass), or if the sensor fails to recognize the valve
position, the error message is generated.
Probable cause Suggested actions
1
Defective rotor seal/stator head.
Exchange the rotor seal/stator head.
2
Defective injection valve.
Exchange the injection valve.

Table of Contents