42 Agilent 1260 Infinity II Preparative Autosampler User Manual
3
Using the Module
Install the External Tray
Install the External Tray
Parts required p/n Description
G7129-60000 External Tray for 5 x 2 mL Vials
G1313-27302 Disposal tube
Preparations • Turn the sampler off
• Open the doors
• Remove the samples
Keep foam and plastic cover in a safe place.
For best temperature performance, and if the external tray is not in use, it is best to cover
the opening for the external tray with the original parts.
1
Lift the foam with a flathead screwdriver.
2
Remove the foam.