Agilent 1260 Infinity II Preparative Autosampler User Manual 141
Error Information
8
Sampler Error Messages
Initialization with Vial
Error ID: 34372
The autosampler attempted to initialize with a vial still in the gripper.
During initialization, the autosampler checks correct operation of the gripper
by closing and opening the gripper fingers while monitoring the motor
encoder. If a vial is still in the gripper when initialization is started, the
gripper fingers cannot close causing the error message to be generated.
Vial in Gripper
Error ID: 34308
The gripper arm attempted to move with a vial still in the gripper.
During specific stages of the sampling sequence, no vial should be held by the
gripper. The autosampler checks if a sample vial is stuck in the gripper by
closing and opening the gripper fingers while monitoring the motor encoder. If
the gripper fingers are unable to close, the error message is generated.
Probable cause Suggested actions
1
Vial still in gripper.
Remove the vial using the Release Vial
function in the user interface. Reinitialize the
autosampler.
Probable cause Suggested actions
1
Vial still in gripper.
Remove the vial using the Release Vial
function in the user interface. Reinitialize the
autosampler.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals