52 Agilent 1260 Infinity II Preparative Autosampler User Manual
3
Using the Module
Transporting the Sampler with a Sample Cooler Installed
Transporting the Sampler with a Sample Cooler Installed
NOTE
When moving the sampler around the laboratory, make sure that any condensed water
inside the thermostat is removed.
Remove the drainage and place a beaker underneath the drain outlet of the sample
cooler. Then carefully tilt the module to the back so that the water inside the
thermostat can safely flow into the leak funnel. If condensate removal is done
improperly, you can harm the electronic of the module.
Otherwise no special precautions are needed for the modules.
WARNING
Heavy weight
The module is heavy.
Carry the module at least with 2 people.
Avoid back strain or injury by following all precautions for lifting heavy objects.
Ensure that the load is as close to your body as possible.
Ensure that you can cope with the weight of your load.
NOTE
Transporting the sampler with a sample cooler installed is only allowed for short
distances. For longer distances, you must separate the units and send them independently.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals