Agilent 1260 Infinity II Preparative Autosampler User Manual 53
Using the Module
3
Transporting the Sampler with a Sample Cooler Installed
If the sampler with a sample cooler needs to be shipped to another location via
carrier, ensure:
The two modules are shipped in separate boxes.
The gripper arm of the sampler is parked, see Park Arm in Agilent Lab
Advisor online help for more information. Also ensure there is no vial in the
gripper arm.
Install the transport protection.
Remove all vials from the drawers.
The condensed water inside of the sample cooler is removed.
CAUTION
Mechanical damage of the module
If the transport assembly is not parked and not protected by the transport foam, the
module could be damaged due to excessive shock of the shipping container during
transport.
Always park the transport assembly before shipment.
Protect the instrument by Transport-Protection (G7129-40050)
Store the installation foam in a save place, to use it for later transport of the
module.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals