Agilent 1260 Infinity II Preparative Autosampler User Manual 85
Preparing the Module
5
Flow Connections to the sampler
Flow Connections to the sampler
1 Place a needle wash solvent reservoir into the solvent cabinet.
2 Connect the Needle Wash Bottle Head Assembly to the solvent reservoir and
close the bottle.
3 Guide the tube of the Needle Wash Bottle Head Assembly through the cover
opening and connect it to the peristaltic pump.
4 Route the drainage of the washport outlet to the waste container.
5 Prime the wash tubing.
6 Install the capillary from the pump outlet into the port 1 of the injection
valve.
Preparations Module is installed in the system.
Use an appropriate solvent based on the sample and mobile phase chemistries.
The composition of the wash solvent should be the most solubilizing compatible solvent (your
strongest diluent). Selecting the wash solvent is part of the method development.
A mixture of 50 % up to 100 % organic solvent in distilled water is a good choice for many
applications.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals