Agilent 1260 Infinity II Preparative Autosampler User Manual 29
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Table 1 Physical Specifications
Type Specification Comments
Weight 19 kg (41.9 lbs) w/o sample cooler
Dimensions
(height × width × depth)
320 x 396 x 468 mm
(12.8 x 15.6 x 18.4 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 350 VA / 350 W / 1195 BTU/h
Ambient operating
temperature
4 - 40 °C (39 - 104 °F), without chiller up
to 55 °C (131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
1
1
If a sample cooler is included the upper value for humidity can be reduced. Please check your lab
conditions to stay beyond dew point values for non–condensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Boiling point ≥56 °C
Auto-ignition temperature ≥200 °C