Agilent 1260 Infinity II Preparative Autosampler User Manual 117
Troubleshooting and Diagnostics
7
Maintenance Functions
Automatic Referencing
Automatic Referencing is required to compensate for small deviations in
positioning of the gripper which may occur after disassembling the module for
repair.
The Automatic Referencing procedure requires Calibration Tool
(G7129-60014) and uses several tray positions as reference points. Because
the tray is a rectangle, a four-point alignment is sufficient to correct all other
vial positions within the tray. On completion of the procedure, the corrected
gripper positions are stored in the instrument firmware.
Start
Moves the transport assembly and gripper arm to the position required to
install the calibration tools.
Next
Mount the calibration fixture. Position P1-C6 and P2-F6.
Next
Mount a 2 mL vial.
Next
Perform autoreferencing.
Next
Remove calibration fixture.
Back
Completes the procedure.
NOTE
Please check help for further explanation.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals