Agilent 1260 Infinity II Preparative Autosampler User Manual 83
Preparing the Module
5
Swage Fittings
Swage Fittings
1
Select a nut that is long enough for the fitting you'll be
using.
2
Slide the nut over the end of the tubing.
3
Carefully slide the ferrule components on after the nut
and then finger-tighten the assembly while ensuring that
the tubing is completely seated in the bottom of the end
fitting.
4
Use a column or injection valve to gently tighten the
fitting which forces the ferrule to seat onto the tubing.
NOTE
Don't overtighten. Overtightening will shorten the
lifetime of the fitting.