Agilent 1260 Infinity II Preparative Autosampler User Manual 165
Maintenance
9
Exchange the Sample Loop Assembly
3
Unscrew the safety cover screw.
Sharp needle
Uncovered needles may cause injuries
➔
Do not touch the tip of the needle.
4
Remove the safety cover.
5
Check if the needle is in maintenance position.
6
Disconnect the loop capillary from the metering device
and from the needle.