168 Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Rotor Seal
1
Open the front door.
2
Remove all capillaries from the injection valve with a
1/4 inch wrench.
3
Loosen each fixing bolt two turns at a time. Remove the
bolts from the head.
4
Remove the stator head and stator ring.