176 Agilent 1260 Infinity II Preparative Autosampler User Manual
9
Maintenance
Exchange the Metering Seal and Piston
15
Reinstall the capillaries.
16
In the Tools section of the Agilent Lab Advisor software
system screen exit Service & Diagnostics >
Maintenance Positions > Change Metering Device:
click Next and wait until the metering device is in Home
position.
17
Select Back to finish the procedure.
18
Close the doors.
19
Perform a pressure test.