Agilent 1260 Infinity II Preparative Autosampler User Manual 33
Site Requirements and Specifications
2
Specifications of the Sample Cooler
General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
➔ The breakdown of the sample cooler unit must be carried out by specialist
refrigeration company.
➔ All media must be disposed of in accordance with national and local regulations.
➔ Please contact your local Agilent Service Center in regard to safe environmental
disposal of the appliance or check www.agilent.com for more info.
Table 4 Performance Specifications Agilent 1290 Sample Cooler
Type Specifications
Operating principle High performance, low-energy consumption
micro-compressor based cooler with
ozone-friendly HFC-134a coolant (42 g),
user-upgradable.
Temperature range from 4 °C to 5 °C below ambient
Temperature settable from 4 – 40 °C in 1 ° increments
Temperature accuracy (<25 °C, <50 % r.H.) 2 °C to 6 °C at a setpoint of 4 °C