Installation of the Optional Sample Cooler
4
Install the Sample Cooler
For the sample cooler installation in a sampler the serial number of the sample cooler
must be DEBAT02001 or higher.
For optimal cooling performance of the autosampler all drawers have to be installed.
1
Ensure that the power switch on the front of the module
is OFF (switch stands out).
2
Ensure that the power cable is removed from the
instrument.
3
Open the 4 screws on the rear of the module.
4
Remove the sample cooler mainframe cover.