Agilent 1260 Infinity II Preparative Autosampler User Manual 81
Preparing the Module
5
Solvent Information
TFE/PDD copolymer tubings, which are used in all Agilent degassers except
1322A, are not compatible with fluorinated solvents like Freon, Fluorinert, or
Vertrel. They have limited life time in the presence of Hexafluoroisopropanol
(HFIP). To ensure the longest possible life with HFIP, it is best to dedicate a
particular chamber to this solvent, not to switch solvents, and not to let dry
out the chamber. For optimizing the life of the pressure sensor, do not leave
HFIP in the chamber when the unit is off.
The tubing of the leak sensor is made of PVDF (polyvinylidene fluoride), which
is incompatible to the solvent DMF (dimethyl formamide).
Sapphire, Ruby and Al
2
O
3
-based ceramics
Sapphire, ruby and ceramics based on aluminum oxide Al
2
O
3
are inert to
almost all common acids, bases and solvents. There are no documented
incompatibilities for HPLC applications.

Table of Contents

Other manuals for Agilent Technologies G7166A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7166A and is the answer not in the manual?

Agilent Technologies G7166A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7166A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals