Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly Level Problems
1-31
A3 RF Assembly Troubleshooting
The self test routine is specifically designed to troubleshoot the A3 RF assembly. Troubleshooting the A3 RF assembly relies on the ability
of the signal generator to complete the associated self test.
1. Run a complete self test or A3 RF assembly specific self tests.
Press:
Utility > Instrument Info > Self Test > Run Complete Self Test
If any self test fails, refer to step 4 of “Running Self Test”.
2. If a complete self test cannot be run, monitor and troubleshoot the “Boot-up Routine” on page 1-20.