Functions and Measurements 3
N9342C/43C/44C User’s Guide 43
Making Distortion Measurements
This section provides information on measuring 
and identifying signal distortion.
Identifying Analyzer Generated Distortion
High level input signals may cause analyzer 
distortion products that could mask the real 
distortion present on the measured signal. Use 
trace and the RF attenuator to determine which 
signals, if any, may be internally generated 
distortion products.
In this example, a signal from a signal generator is 
used to determine whether the harmonic distortion 
products are generated by the analyzer.
1 Input a signal (200 MHz, –10 dBm) to the ana-
lyzer  RF IN connector.
2 Set the analyzer center frequency and span:
• Press [Preset] (factory preset)
• Press [Freq] > 400 > {MHz}
• Press [Span] > 700 > {MHz}
The signal produces harmonic distortion products 
(spaced
 200  MHz  from  the original 200 MHz signal). 
Figure 3-13  Harmonic distortion