2 Getting Started
24 N9342C/43C/44C User’s Guide
Adding an Option
Pressing  [System] > {More} > {Service} > {Add Option} 
brings up a dialog box for entering the option 
license code. Use the numeric keypad to input the 
option license code and then use the [ENTER] key as 
a terminator. If the analyzer recognizes the option 
license code, a message “Option activated 
successfully” will appear in the status line. 
Otherwise, a message “Invalid option licence” will 
appear in the status line. Press [System] > {System 
Info} > {Installed Options} to view the options.
Show Error
Pressing [System] > {System Info} > {Error history} 
accesses a list of the 30 most recent error 
messages. The most recent error will appear at the 
bottom of the list. If the error list is longer than 30 
entries, the analyzer reports an error message 
[–350, Query overflow]. For more information, 
refer to “Error Messages" on page 143.
Perform Calibration
The N934xC provides three manual calibration 
function to calibrate the time base and amplitude. 
The analyzer should warm up for 30 minutes 
before calibration. 
Time Base Calibration
Perform a time base calibration to guarantee the 
frequency accuracy.
When the calibration function is triggered, the 
current measurement is interrupted and a gauge 
displays on the LCD. A message will display on the 
LCD which indicates the calibration is finished, 
and the interrupted measurement restarts.
Please refer to the operation procedures below:
1 Input a 10 MHz, 0 dBn signal to EXT TRIG IN.
2 Press  [System] > {More} > {Service} > {Calibration} > 
{Time Base by Ext} to initiate a calibration.