Functions and Measurements 3
N9342C/43C/44C User’s Guide 77
Calibrate with a T-combo calibrator
Using a T- combo calibrator is a three step process 
requiring the user to manually change the 
open- short- load components. The default order of 
the calibration process is open- short- load.
1 Press  [Freq] > {Calibrate} to bring up the dialog 
box for calibration instruction. Follow the 
on- screen instruction to connect the Open 
component of the T- combo calibrator to the RF 
OUT connector of the analyzer. Press [ENTER] to 
start the open calibration.
2 Follow the instruction to connect the short 
component of the T- combo calibrator to the RF 
OUT connector of the tester. Press [ENTER] to 
start the short calibration.
3 Connect the Load component of the T- combo 
calibrator to the RF OUT connector of the tester. 
Press  [ENTER] to start the load calibration.
A  Calibrated indicator is shown on the upper left 
corner of the screen when the calibration is 
completed. Then, the analyzer is ready for your 
further cable measurement.
Measuring Cable Reflection
Press  [Mode] > {Cable&Antenna Test} > {Reflection 
Measurement} to select the Reflection Measurement 
Mode. It allows you to set the Y axis of the 
graticule as return loss, VSWR or reflection 
coefficient value. For further information on 
reflection measurement, please refer to “Making a 
Reflection Calibration" on page 53.
Measuring Distance to Fault 
For this measurement, an internally generated 
signal emits from the RF OUT port of the tester 
and is transmitted to the cable and antenna system 
(DUT). The tester receives and processes the signal 
power reflected from faults and imperfections. In 
performing a DTF measurement, the tester uses