Chapter 4
Source Key and Programming Commands
External Source Configuration
48
Source Key and Programming
Commands
4.1.2 Show Setup
Pressing this immediate action key displays a form showing you the current configuration of the external
source at the specified IP address.
4.1.3 Model Number
Information parameter which displays the model number of the signal source connected at the specified IP
Address. The model number displayed will be determined by sending a “*IDN?” to the entered IP address.
The model number is the second field of the “*IDN?” returned value (Format of the *IDN? string is
‘<Company Name>, <Model Number>, <Serial Number>, <Firmware Revision>’).
Mode: Spectrum Analysis
Key Path:
Source, Ext Src Config
Dependencies and
Couplings:
Parameter is disabled if active mode is not Spectrum Analyzer
or a connection has not been established at the specified IP
address.
Remote Command
Notes:
No SCPI. Front Panel only.
Mode: Spectrum Analysis
Key Path:
Source, Ext Src Config, Show Setup
Remote Command: :SOURce:EXTernal:MODel?
Factory Preset: ------
State Saved: No
Range: Uppercase, Numeric
Remote Command
Notes:
Query Only
Example: SOUR:EXT:MOD?

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals