Chapter 4
Source Key and Programming Commands
Configure Sweep
57
Source Key and Programming
Commands
4.4.3 Power Sweep
The external source amplitude is swept over the specified number of sweep points.
Mode: Spectrum Analysis
Key Path:
Source, Config Sweep
Remote Command: :SOURce:EXTernal:SWEep:POWer:SPAN <rel_ampl>
:SOURce:EXTernal:SWEep:POWer:SPAN?
:SOURce:EXTernal:POWer:MODE FIXed|SWEep
:SOURce:EXTernal:POWer:MODE?
Factory Preset: 0.00 dB, FIXed
State Saved: Saved in instrument state.
SCPI Resolution Max: 1.00 dB
Dependencies and
Couplings:
Parameter is disabled if active mode is not Spectrum Analyzer
mode or a connection has not been established at you specified
IP address.
Example: SOUR:EXT:SWE:POW:SPAN 10 dB
SOUR:EXT:SWE:POW:MODE SWE