72 Chapter 5
Hints and Tips
FAQs and Helpful Hints and Tips
Hints and Tips
But 7.02 GHz is not in the range of the signal source.
So the offset or stop frequency should be changed until the calculation
is less than the source maximum frequency. (Find the source maximum
frequency using the PSA command
SOURce:EXTernal:FREQuency:STOP? MAX)
Example 6: Offset Causes VALID Negative Source Settings
Using any PSA and any source:
PSA settings:
Start Freq = 2.5 GHz
Stop Freq = 3.0 GHz
Sweep Points = 101
•Offset =
3.1 GHz, On
•Harmonic = OFF
This will not cause the “Freq range of ext src exceeded” error.
Reason:
The signal source Start Frequency and the Stop Frequency are both
negative. In this case:
In this case both the stop and start frequencies are negative, so this
example looks invalid. But in this special case, the PSA can change its
internal mixing to use a negative mixing harmonic. Then the PSA sets
the source start frequency to the positive value of the calculated stop
frequency. It also sets the source stop frequency to the positive value of
the calculated start frequency. This causes the mixing products to
sweep in reverse, and allows the measurement to be made.
Source Stop Freq
12E 9() 5E 9()()+
12E 9() 10E 9()
101 1
------------------------------------------
⎝⎠
⎛⎞
+
1
------------------------------------------------------------------------------------------------------ 7.02E 9() = 7.02 GHz==
Source Start Freq
2.5E 9() 3.1E 9()()+
3.0E 9() 2.5E 9()
101 1
---------------------------------------------
⎝⎠
⎛⎞
1
--------------------------------------------------------------------------------------------------------------- 0.605E 9() = –605 MHz==
Source Stop Freq
3.0E 9() 3.1E 9()()+
3.0E 9() 2.5E 9()
101 1
---------------------------------------------
⎝⎠
⎛⎞
+
1
--------------------------------------------------------------------------------------------------------------- 0.095E 9() = –95 MHz==

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals