Chapter 4
Source Key and Programming Commands
Configure Sweep
54
Source Key and Programming
Commands
4.4 Configure Sweep
4.4.1 Offset Sweep
Offset sweep offsets the frequency of the signal generator from the frequency of the PSA. One application
for offset sweep is testing mixer conversion loss. See the following graphic.
Mode: Spectrum Analysis
Key Path:
Source, Config Sweep
Remote Command: :SOURce:EXTernal:SWEep:OFFSet:FREQuency <freq>
:SOURce:EXTernal:SWEep:OFFSet:FREQuency?
:SOURce:EXTernal:SWEep:OFFSet[:STATe] ON|OFF|1|0
:SOURce:EXTernal:SWEep:OFFSet[:STATe]?
Factory Preset: 0.00 Hz, OFF
State Saved: Saved in instrument state.
SCPI Resolution Max: 1 Hz