Chapter 4
Source Key and Programming Commands
Normalize
58
Source Key and Programming
Commands
4.5 Normalize
4.5.1 Store Reference Trace
Copies trace 1 into trace 3.
Mode: Spectrum Analysis
Key Path:
Source, Normalize
Remote Command: :TRACe:COPY TRACE1, TRACE3
Dependencies and
Couplings:
Parameter is disabled if active mode is not Spectrum Analyzer
or a connection has not been established at the specified IP
address.
A reference trace should be stored prior to turning on trace
normalization either through pressing the Store Ref Trace
softkey or completing an Open/Short calibration.
The stored reference trace (trace 3) is cleared when key is
pressed. If you wish to re-use this calibration data you should
save the trace (File > Save > Type (Trace) > Save Now) and then
recall it following completion of the preset.
Example: TRAC:COPY TRACE1, TRACE3