68 Chapter 5
Hints and Tips
FAQs and Helpful Hints and Tips
Hints and Tips
Examples Exceeding the Source Frequency Range or
the Spectrum Analyzer Frequency Band.
These examples will all result in an instrument error until the
measurement settings are changed.
The following reference information applies to all of the examples.
Example 1: Bad Source Start Frequency
Using any PSA and any source:
PSA settings:
Start Freq = 0 Hz
Stop Freq = 500 MHz
Sweep Points = 101
•Offset = OFF
•Harmonic = OFF
This will cause the “Freq range of ext src exceeded” error.
Reason:
The signal source must be able to sweep one “bucket’s” worth of
frequency range before the requested PSA start frequency. In this case:
But
5 MHz is not in the range of the signal source.
Source Start Freq
PSA Start Freq Offset Freq+
PSA Stop Freq PSA Start Freq
Sweep Points 1
------------------------------------------------------------------------------
⎝⎠
⎛⎞
Harmonic Number
----------------------------------------------------------------------------------------------------------------------------------------------------------------=
Source Stop Freq
PSA Stop Freq Offset Freq+
PSA Stop Freq PSA Start Freq
Sweep Points 1
------------------------------------------------------------------------------
⎝⎠
⎛⎞
+
Harmonic Number
----------------------------------------------------------------------------------------------------------------------------------------------------------------=
Source Start Freq
00+
500E 6() 0
101 1
------------------------------
⎝⎠
⎛⎞
1
------------------------------------------------------5E6() = –5 MHz==
Source Stop Freq
500E 6() 0+
500E 6() 0
101 1
------------------------------
⎝⎠
⎛⎞
+
1
------------------------------------------------------------------------ 5 0 5 E 6() = 5.05 MHz==

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals