Chapter 4
Source Key and Programming Commands
Amplitude
52
Source Key and Programming
Commands
Dependencies and
Couplings:
Parameter is disabled if active mode is not Spectrum Analyzer or a
connection has not been established at the specified IP address.
Setting state to ON initializes the external source based on the
current settings of the spectrum analyzer. The external source
display will be turned OFF and the following external source
parameters will be initialized:
Start/Stop Frequency
Start/Stop Amplitude
•Number of Points
Dwell Time
Sweep type set to stepped
In addition setting the external source control state to ON also sets
the following on the PSA:
Video Bandwidth = 50 MHz, the PSA video bandwidth is fixed
at 50MHz while external source control state is set to ON.
Detector Type to Sample
Trigger Polarity
Trigger Source to External Front
Gate Polarity to Positive
Gate Trigger to External Rear
Gate Delay
•Gate Length
•Sweep Time
Resolution Bandwidth
Gate State to ON
If the calculated external source start or stop frequency exceeds the
frequency range of the external source the sweep is stopped until
you adjust either analyzer start or stop frequency, the frequency
offset (if ON) or harmonic (if ON) such that the frequency range
can be swept by the signal source and PSA simultaneously. Once
the frequency range of the external source is valid, the sweep
resumes.
Example: SOUR:EXT:POW
22 dBm
SOUR:EXT:POW ON

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals