TriScroll 600 Dry Scroll Vacuum Pump
D R A F T 4 / 2 9 / 0 4
1
Major Maintenance
General Information
Agilent TriScroll 600 series pumps are designed to provide years of trouble-free service if maintenance
procedures and intervals are observed. Bearing grease replenishment and tip seal replacement are
recommended when the pump base pressure has risen to an unacceptably high level for your
application. Bearings, rotary seals and O-rings should also be replaced if the pump exhibits humming
or grinding noises from the bearings. Main bearing life may be shortened if your application requires
the pumping of high quantities of water vapor. Use of the bearing purge will keep this water from
impacting bearing life.
Required Equipment
Maintenance Tool Kit: PTSS0600TK (page 2)
Major Maintenance Kit: PTSS0600MK (page 4)
Arbor Press: 1/2 ton or larger, 10" work diameter capacity, 8" capacity over table
Oven: 400
o
F temperature capability, 11" wide x 11" deep x 4" high minimum chamber,
500 watt or higher heating capacity
Heat Resistant Surface
Vacuum Gauge: Capable of measuring pressure of 5 mTorr to 20 mTorr with an accuracy of
± 1 mTorr. A capacitance manometer or Pirani gauge is recommended.

Other manuals for Agilent Technologies TriScroll 600 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies TriScroll 600 Series and is the answer not in the manual?

Agilent Technologies TriScroll 600 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTriScroll 600 Series
CategoryWater Pump
LanguageEnglish

Related product manuals