Agilent Vapor Generation Accessory VGA 77 and VGA 77P User’s Guide 15
Condition Altitude Temp (°C)
Humidity (%RH)
non-condensing
Non-operating (transport) 0-2133 m
(0-7000 ft)
5-45 20-80
Operating within performance
specifications
0-853 m (0-2800 ft)
853-2133m (2800-7000 ft)
10-35
10-25
8-80
Power
If necessary, replace the power cord only with a cord equivalent to
the one specified below.
Information is correct at time of printing—information on the back of
the accessory is the most up-to-date.
Voltage 100, 120, 220, or 240 volts AC
Frequency 50-60 Hz
Consumption approx. 20 VA
Connections
Mains inlet coupler: 6 A 250 VAC IEC type
Mains power cord connector:
Australia 10 A 250 VAC Complies with AS3112
USA 10 A 125 VAC Complies with NEMA 5-15P
Europe 6 A 250 VAC Complies with CEE7 sheet vii or NFC61.303VA
Gas Supplies
Argon or nitrogen, 99.99% pure (minimum)
NOTE
Only argon gas can be used when the VGA 77 is coupled with an ICP
spectrometer.
The inert gas must be dry and dust-free, otherwise you must insert a
suitable filter in the supply line.