Getting Started
24 Agilent Vapor Generation Accessory VGA 77 and VGA 77P User’s Guide
Installing the Pump Unit
WARNING
Moving parts
The peristaltic pump has moving parts that can cause damage to hands or
fingers if caught underneath the clamps during operation. Keep hands away
from the peristaltic pump while it is running.
To install the VGA 77 on an Agilent AA:
Insert the hooks on the back of the unit into the slots located on
either side of the Agilent AA sample compartment.
To install the VGA 77 on an Agilent 700 Series ICP-OES
instrument:
1 Insert the four stainless steel thumbscrews into the four holes
towards the front of the instrument sample compartment (two on
each side).
2 Hook the four grooves in the VGA 77 mounting bracket over the
four thumbscrews in the sample compartment.
For an instrument other than an Agilent AA or Agilent 700 Series
ICP-OES, you should place the VGA 77 flat on a bench, close to the
sample compartment of the instrument.
When the VGA 77 is in position, fit the drip tray to the front of the
pump unit by positioning the large holes over the studs on the front
of the VGA 77 and sliding the tray downwards.
Installing a Reagent Module
To install a reagent module into the VGA 77:
1 Loosen the locking screw at the lower rear left hand corner of the
pump unit.
2 Lower the reagent module into the space adjacent to the left side
of the pump unit, ensuring the reagent module locating studs are
fully home in the slots.
3 Wind the locking screw in until it holds the module firmly in
place.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies VGA 77 and is the answer not in the manual?

Agilent Technologies VGA 77 Specifications

General IconGeneral
BrandAgilent Technologies
ModelVGA 77
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals