BYK-Gardner GmbH Configuration | 8
301 200 085 E 2309 wave-scan 3 / wave-scan 3 dual 49
You have following options:
n Sample: Defines how many measurements are to be taken.
Column 1 .. 3: For each column you can decide, which value is to be dis-
played.
The statistic function comprise following data:
Off: Column w/o data
Min: Minimum value
Max: Maximum value
Stdev: Standard deviation
Range: Measurement interval
Value: Actual value
Mean: Arithmetic average
NOTICE
To switch off statistics you must set n = 1. To save this setting it is required, to set
all columns to Off.
8.1.3 Scan Length
You can define the distance the device is to be moved across the surface of the
sample in order to complete a measurement.
You have following options:
5cm
10 cm
20 cm
NOTICE
The selected scan length has impact on measurement accuracy. Using a short scan
length will result in strong statistical skewing of measurement values for large
wavelengths. Thus, when using the Wd scale with a scan length of 5 cm, at least 3
measurements should be made per sample to ensure representative values.

Table of Contents

Other manuals for Altana BYK wave-scan 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Altana BYK wave-scan 3 and is the answer not in the manual?

Altana BYK wave-scan 3 Specifications

General IconGeneral
BrandAltana
ModelBYK wave-scan 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals