BYK-Gardner GmbH Configuration | 8
301 200 085 E 2309 wave-scan 3 / wave-scan 3 dual 49
You have following options:
• n Sample: Defines how many measurements are to be taken.
• Column 1 .. 3: For each column you can decide, which value is to be dis-
played.
The statistic function comprise following data:
• Off: Column w/o data
• Min: Minimum value
• Max: Maximum value
• Stdev: Standard deviation
• Range: Measurement interval
• Value: Actual value
• Mean: Arithmetic average
NOTICE
To switch off statistics you must set n = 1. To save this setting it is required, to set
all columns to Off.
8.1.3 Scan Length
You can define the distance the device is to be moved across the surface of the
sample in order to complete a measurement.
You have following options:
• 5cm
• 10 cm
• 20 cm
NOTICE
The selected scan length has impact on measurement accuracy. Using a short scan
length will result in strong statistical skewing of measurement values for large
wavelengths. Thus, when using the Wd scale with a scan length of 5 cm, at least 3
measurements should be made per sample to ensure representative values.