8 | Configuration BYK-Gardner GmbH
56 wave-scan 3 / wave-scan 3 dual 301 200 085 E 2309
Now you can use the function Standard Measurements [}29] to check the toler-
ances. Details see Testing the Instrument [}24].
If the differences are within your tolerances, you can continue to measure with
your instrument. Otherwise clean the standards and repeat, see Cleaning Intruc-
tions [}60].
NOTICE
All measurements on the checking tiles are stored in the instrument. Transfer the
measurement data to “smart-chart” to document all test case performed.
8.6 Factory Reset
If you are having technical problems with your instrument, you can perform a fall-
back to the factory settings.
Confirm the security message to perform the fallback. You must enter the pass-
word “byk-instruments”; see also section Technical Data [}68] > General Data.
After fallback all your personal configuration and measurement data in the device
will be lost. The instrument starts in a fresh state.