Portable Flue Gas Analyser
User Guide
Lancom 4
15 - 4
Step Function See Insight
User Guide
Page
16 Complete the Edit Trigger dialog box as shown in the exam-
ple.
17
Select the Save icon from the Toolbar
47
18 Enter the project name and store in the desired directory us-
ing the Save dialog box.
19
Select Activate Trigger
in the Record menu.
Having completed the steps above the system will start recording when the
SO
2
level exceeds 300 ppm. Once the condition is met the acquisition can be
halted at any time using the icon on the Toolbar.
Data can be examined in either the Base Data Table or the User Table. Once
acquisition is stopped additional columns of data can be added to the User
Table. These columns will be populated from data stored in the Base Data
table. A graph can be plotted by selecting the
icon on the Toolbar.