PN 305725901, Rev YF
REG DEFINITION UNIT SCALING ACCESS
28 Serial communication mode (SerialMode= 26 Hex) 1 r,w
29 S lave address (NodeAddress= 27 Hex) 1 r,w
30 Baud rate (Baud= 25 Hex) 1 r,w
31 Modbus protocol setting (ParityAndStop= 92 Hex) 1 r,w
32 Counts to voltage coecient (AD16Oset= 01 Hex) 1 r,w
33 Sensor assembly setting (RefPeriod= 04Hex) second 1 r,w
34 Sensor assembly setting (SmplPeriod= 05 Hex) second 1 r,w
35 Sensor temperature coecient (CellTempSpan=06 Hex) 100 r,w
36 Sensor temperature coecient (CellTempOset=07 Hex) 10 r,w
37 Process pressure coecient (CustomerSpan= 08 Hex) 1000 r,w
38 Process pressure coecient (CustomerOset= 09 Hex) 100 r,w
39 Sensor pressure lter coecient (Pressure1Filter= 0A Hex) 100 r,w
40 Flow lter coecient (Flow Filter= 0B Hex) 100 r,w
41 Sensor temperature lter coecient
(CellTempFilter= 0C Hex)
100 r,w
42 Process pressure lter coecient (CustomerFilter= 0D
Hex)
100 r,w
43 Current output conversion parameter
(MaOutSpan= 0E Hex)
1 r,w
44 Current output correction parameter
(MaOutO set= 0F Hex)
10 r,w
45 Current output coecient (AnalogOutSpan= 10 Hex) 1 r,w
46 Current output coecient (AnalogOutOset= 11 Hex) 1 r,w
47 Process pressure coecient (Process Span= 12 Hex) 10 r,w
48 Process pressure coecient (Process Oset= 13 Hex) 10 r,w
49 Flow control parameter (ProBandLoop1= 16 Hex) 1 r,w
50 Flow control parameter (TsLoop1= 17 Hex) 1 r,w
51 Flow control parameter (TiLoop1= 18 Hex) 1 r,w
52 Flow control parameter (SetPointLoop1= 19 Hex) 1 r,w
53 Flow control parameter (ActuatorLoop1= 1A Hex) 1 r,w
54 Flow control parameter (uMaxLoop1= 1B Hex) 10 r,w
55 Flow control parameter (uMinLoop1= 1C Hex) 10
56 Heater control parameter (proBandLoop2= 1D Hex) 1 r,w
57 Heater control parameter (TsLoop2= 1E Hex) 1 r,w
58 Heater control parameter (T iLoop2= 1F Hex) 1 r,w
59 Heater control parameter (SetPointLoop2= 20 Hex) 1 r,w
60 Heater control parameter (ActuatorLoop2= 21 Hex) 1 r,w
61 Heater control parameter (uMaxLoop2= 22 Hex) 1 r,w
62 Heater control parameter (uMinLoop2= 23 Hex 1 r,w
63 Gas based coecient (Ppm W = 24 Hex) 1000 r. w
64 Flow control parameter (TempLoopDelay= 29 Hex) second 1 r,w
6-6 | 3050-DO Moisture Analyzer