PN 305725901, Rev YF
REG DEFINITION UNIT SCALING ACCESS
65 Unit of measurement selection (MoistureUnits= 2B Hex) 1 r,w
66 Concentration alarm range (Alarm 1H i= 2C Hex) 1 r,w
67 Concentration alarm range (Alarm 1Lo= 2D Hex) 1 r,w
68 Concentration alarm range (Alarm 2Hi= 2E Hex) 1 r,w
69 Concentration alarm range (Alarm 2Lo= 2F Hex) 1 r,w
70 Process pressure units (PressUnits= 30 Hex) 1 r,w
71 Bypass valve state (Bypass State= 31 Hex) 1 r,w
72 Zero coecient (FrequencyOset= 38 Hex) Hz 1000 r,w
73 Gas based ow coecient (Flow Coe0= 39 Hex) 1000 r,w
74 Gas based ow coecient (Flow Coe1= 3A Hex) 1000 r,w
75 Process pressure upper limit (Flow Coe2= 3B Hex) kPa 1 r,w
76 Flow meter parameter (FlowSpan= 3C Hex) 100 r,w
77 Auto verication hour (CalHour= 3D Hex) 1 r,w
78 Verication time in minutes (CalPeriod= 3E Hex) minute 1 r,w
79 Auto verication day of month (CalMonthDays= 3F Hex) 1 r,w
80 Auto verication day of week (CalWeekDays= 40 Hex) 1 r,w
81 Hourly, daily, monthly auto verication (CalType= 41 Hex) 1 r,w
82 Software revision number (Version= 47 Hex) 100 r,w
83 Reset variable (System Reset= 48 Hex) 1 r,w
84 Default memory ag (new Flag= 49 Hex) 1 r,w
85 Read only except test mode (AnalogInput= 4E Hex) 10 r,w
86 Read only except test mode (AnalogOutput= 4F Hex 1 r,w
87 Used in test m ode (ReadTestFixture= 50 Hex) 1 r
88 Sample = 0,generator = 1,reference = 2
(ValveState= 56 Hex)
1 r,w
89 Current hour (Hour= 57 Hex) hour 1 r,w
85 Read only except test mode (AnalogInput= 4E Hex) 10 r,w
86 Read only except test mode (AnalogOutput= 4F Hex 1 r,w
87 Used in test mode (ReadTestFixture= 50 Hex) 1 r
88 Sample = 0,generator = 1,reference = 2
(ValveState= 56 Hex)
1 r,w
94 Current year (Year= 5C Hex) 1 r,w
95 Used in test only (WriteTestFixture= 63 Hex) 1
96 Current day of week (WeekDay= 6D Hex) 1 r,w
97 Verication coecient (MoistureSpan= 6E Hex) 1000 r,w
98 Track or hold current output ag (HoldOut= 71 Hex) 1 r,w
99 Fast cycle = 0, slow cycle= 1 (SlowTiming= 72 Hex) 1 r,w
100 Minimum verication duration
(VerDurationLim it= 73 Hex)
minute 1 r,w
101 Adjust span in verication ag (AdjustSpan= 74 Hex) 1 r,w
102 Alarm 1 enable ag (EnableAlarm 1= 75 Hex) 1 r,w
Serial Communication Interface | 6-7