PN 305725901, Rev YF
REG DEFINITION UNIT SCALING ACCESS
230 LowSpan Limit (LowSpanLimit= 95 Hex) 100 r,w
231 EquilCal (EquilCal = 7D Hex) 1 r,w
232 (EquilZero = 7E H EquilZero Hex) 1 r,w
233 ZeroDurationLim it (ZeroDurationLimit = 7F) 1 r,w
234 ZeroCycleCount (ZeroCycleCount = 80) 1 r,w
235 ZeroHour (ZeroHour = 81) 1 r,w
236 ZeroPeriod (ZeroPeriod = 82 ) 1 r,w
237 ZeroMonthDays (ZeroMonthDays = 83) 1 r,w
238 ZeroWeekDays (ZeroWeekDays = 84) 1 r,w
239 ZeroType (ZeroType = 85) 1 r,w
240 MaxZeroError (MaxZeroError = 86) Hz 100 r,w
241 MaxZeroDrift (MaxZeroDrift = 87) Hz 100 r,w
242 MoistureSpanSlow (MoistureSpanSlow = 88) 1000 r,w
243 FreqOsetSlow (FreqOsetSlow = 89) 1000 r,w
Register #0 holds the moisture concentration value not held during Verica-
tion. This value is in PPM.
A pair of registers #135 and #136 provide the same moisture concentration
value in oating point Modicon standard.
Register #1 holds the moisture concentration value held during verication
cycle. The units of measurement are changing depending of ag status located
in register #65.
A pair of registers #137 and #138 provide the same information in oating
point format.
Register #11 is a DataState register, which is designed to synchronize the data
acquisition process. This ag is cleared by read. A value of one indicates new
data is available.
Register #8 is the SystemState variable, which is set to alarms and warnings.
This value is decoded according to the table below.
Serial Communication Interface | 6-9