Controller / User Interface | 4-63
Serial Number Update
Used to re-enter the analyzer’s serial number (to correspond with
the analyzer data) after replacing the analyzer’s embedded software
(EPROMs) or EEPROM (see also “Restoring Analyzer Configuration
Parameters” in this chapter).
Model ID
Displays a code (see Analyzer Legend, below) that the software uses
to define the AMETEK analyzer model the software is currently com-
municating with (not user adjustable). Analyzer models include:
151 = M931S (Model 931S Analyzer)
152 = M932S (Model 932S Analyzer)
153 = M933S (Model 933S Analyzer)
Display
During normal operation, displays the species being measured (top
line) and its real-time status (bottom line).
When an Auto-Setup is in progress, a message indicating the current
stage of the Auto-Setup is displayed on the top line (the concentration
value continues to be displayed on the bottom line).
During normal conditions, the text is Green; when there is a Warning
or Fault alarm present, the text is Red. If an alarm is present, refer to
the “Diagnostics and Troubleshooting” section in Chapter 6 for details
about the alarm and corrective action to take to reset the alarm.
Refresh button
Reverts the changes made to this screen back to their original set-
tings, but only before you select the Update button or OK. When you
click Refresh, any values that you have changed but do not wish to
set permanently will revert to the settings that were last saved to the
system’s EEPROM.
HC Reset button
Resets the analyzer’s Host Controller board.