Maintenance and Troubleshooting | 6-27
Observe the top line of the Display readout. The following messages
are displayed during the Auto-Setup sequence:
Adjusting PMT Level
Adjusting Lamp Pulses
Adjusting PMT Balance
Adjusting PMT Level
Adjusting Lamp Pulses
Setting Up ALC
• Off/Completed
When the message “Off/Completed” is displayed, the Auto-Setup is
complete.
4. From the Optical Bench dialog box, check the PMT Level and PMT
Balance values again. The PMT Level should be between 7200–7400
mV, while the PMT Balance should be between 2000 mV and 8000 mV.
5. If further adjustments are required, refer to “Manipulating the PMT
Level and PMT Balance,” following this section.
Cancel Auto-Setup
To cancel the Auto-Setup, click Cancel Auto-Setup on the Optical Bench
dialog box.