Specications | 2-5
Ambient Limits
Temperature
From 0–50 °C (32–122 °F).
For applications using an unheated Measuring Cell, the surrounding
ambient temperature must be at least 5 °C above the sample dew point
temperature. In high dewpoint applications, a heated Measuring Cell and
heat-traced Sample/Vent Lines must be used to prevent condensation of
liquids in the sample system.
Humidity
0–95 % RH
If ambient conditions typically exceed 70 % RH, the environmental purge
options is required.
Note: Environmental purge set at 10 psi will consume a 300 ft
3
cylinder in
approximately 1 year.
Maximum Altitude
2000 m
Zero Gas
UHP nitrogen, or Instrument Air for non-flammable streams.
Sample Transport
Application dependent.
Sample Pressure Requirement
Maximum Cell Pressure 1380 kPag (200 psig)
Without Filterblock Assembly (typical):
Application dependent.
With Filterblock Assembly (optional):
Application dependent.