6-20   |   Model 931S / Model 932S UV Analyzers
10.  Open the Configurator file “M931S.dyc” or “M932S.dyc” (click     on 
the Standard Toolbar to display two Trend graphs, Figure 6-8):
•  One for PMT Msr (Measure) Signal.
•  One for PMT Ref (Reference) Signal.
The inputs are typically assigned to the channels at the factory. If they 
are set up as described in Step a through Step h, continue with Step 
11; if they are not set up, or they need to be modified, follow these 
steps.
  To assign inputs to the channels on the Msr PMT Signals Properties 
and Ref PMT Signals Properties windows:
a.  Right-click on the Msr PMT Signals Trend graph window and 
click Scan from the context menu to view the Msr PMT Signals 
Trend Properties dialog box (Figure 6-9).
b.  To assign an input to a channel click the first (top) button under 
Channels to select it.
NOTE
Figure 6-8. 
Congurator screen 
after opening the 
M93XS.dyc le.