6-50   |   Model 931S / Model 932S UV Analyzers
  Decimal  Alarm
 Bit  Value  Type  Alarm Condition / Description and Suggested Corrective Action
 2  4  Fault  f: Sample gas pressure too low
The sample pressure is below 1 % of the full-scale range.
Note:  This alarm is used only by analyzers equipped with the optional 
pressure transducer. 
Corrective Action:
•  Check the sample gas pressure gauge and ensure that it is at an appropriate 
ow and pressure, as recommended on the system drawings.
•  From the Output Setup dialog box, conrm the sample gas Low Scale and 
High Scale pressure settings are correct.
•  Check for leaks in the sample system (see “Sample System Leak Check” in 
Chapter 3).
•  Check the solenoids (backpan).
If this doesn’t correct the alarm, take appropriate safety precautions, open the 
Lower Enclosure, and:
•  Check the pressure transducer.
•  Replace the Customer I/O board. Contact AMETEK to verify operation 
before removing this board.
 3  8  Fault  f: On-Chip A/D not responding
The Host Controller board’s internal analog-to-digital converter (A/D) is not 
responding.
Corrective Action:
Take appropriate safety precautions, open the Lower Enclosure, and:
•  Replace the Host Controller board. Contact AMETEK to verify operation 
before removing this board.
Setup (tab)Output