26 | IPS-4 Zone1/Analyzer
The surrounding ambient temperature of the analyzer should be main-
tained between -20 °C to +50 °C (-4 °F to +122 °F). If the analyzer system
is mounted within an AMETEK-designed shelter, these conditions are
taken into consideration in the design of shelter.
To mount the analyzer:
Non-Shelter Installations:
In a suitable, accessible location, mount the analyzer on a vibration-free
wall or instrument rack.
Shelter Installations:
In a suitable, accessible location, install the analyzer in a vibration-free
location.
Refer to customer-specific shelter drawings in the “Supplemental In-
formation” tab of this manual or the Documentation Package shipped
with the analyzer.
Outdoor Installations:
In most cases, the analyzer must be shielded from harsh environmental
elements such as wind, rain, and sun. In these cases a rain shelter or sun
shade is required.
Also, proper water drainage must be considered in the mounting location
to allow for suitable and safe working conditions for service personnel.