EasyManuals Logo

Analog Devices MAXREFDES106# User Manual

Analog Devices MAXREFDES106#
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #27 background imageLoading...
Page #27 background image
©2023 Analog Devices, Inc. All rights reserved.
27
MAXREFDES106# Health Sensor Platform 4.0
Associated Tabs
The tabs for configuring the ICG algorithm are the Mode, ECG Channel, ECG Mux, ECG RLD, BioZ
Drive, BioZ Receive, and BioZ Mux tabs. The ICG algorithms operate on BioZ and ECG input data, but
only support 512Hz sampling rate from both ECG and BioZ.
Minimum Settings
The following settings satisfy the minimum requirements for collecting ICG data. The rest of the settings
can be configured further but can also be left at their power-on defaults. The ICG algorithm uses these
default settings.
Mode: Select HR [ECG] + ICG [BioZ] in Quick Configuration window.
Mode: Enter the Distance Between the V+/- Electrodes, Gender, Height, Weight and Body
Build in the Stroke Volume Parameters window.
Figure 31. Mode Tab: Select HR [ECG] + ICG [BioZ] and Enter Stroke Volume Parameters
These settings satisfy the requirements to start an ICG measurement and the rest of the tabs can be left
at their power-on defaults. To improve the accuracy of the measurement, a BioZ calibration should be
done (see BioZ Calibration section for details).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analog Devices MAXREFDES106# and is the answer not in the manual?

Analog Devices MAXREFDES106# Specifications

General IconGeneral
BrandAnalog Devices
ModelMAXREFDES106#
CategoryMedical Equipment
LanguageEnglish