Technical data
qTOWER
3
/ qTOWER
3
G Edition 03.19 17
3 Technical data
Block format
96 wells
Sample volume
10 80 µL
Lid temperature
30 110 °C
Temperature gradient
40°C (qTOWER
3
G only)
Max. heating rate*
max. 8°C/s,
Ø
7°C/s
(depending on consumables used)
Max. cooling rate*
max. 6 °C/s,
Ø
5.5°C/s
(depending on consumables used)
Heating rate adjustment
min. 0.1 °C/s
Temperature uniformity
(15 s after starting the clock)
±0.15 °C at 55 °C
±0.25 °C at 72 °C
±0.50 °C at 95 °C
Temperature range
3 °C 99 °C
Control accuracy
± 0.1 °C
Temperature increments
min. 0.1 °C/cycle
Time increments
min. 1 s/cycle
Heated lid
Manual opening mechanism, automatic contact
pressure
30 kg, automated
Dimensions
(H x W x D)
59 cm x 27.5 cm x 27.5 cm
70 cm x 27.5 cm x 50 cm when opened
Mass
30 kg
Noise level
45 dB
Energy supply
Operating voltage
Line frequency
100 V, 115 V, 230 V
50 60 Hz
Power consumption (max.)
max. 850 W
Instrument fuses
2 x 10 AT / 250 V
Safety circuits
Monitoring of the closing mechanism of the device
(lock with handle), over-temperature protection in
the heated lid
Operating conditions
Air pressure: 0.7 bar to 1.06 bar
Max. permissible altitude: 2000 m
Humidity: max. 70 %
Operating temperature: +15 °C to +35 °C

Table of Contents