10580-00176 104
OTA Scrambling Code Lock
[:SENSe]:OTA:SCODe:LOCK OFF|ON|0|1
[:SENSe]:OTA:SCODe:LOCK?
This command refers to the primary scrambling codes used for W-CDMA over-the-air
measurements.
When scrambling code lock is ON, the instrument will not search for any new scrambling
codes and will only measure the codes that have already been detected.
When scrambling code lock is OFF, the instrument will dynamically update the list of
scrambling codes that are to be included in the current measurements.
This feature has no effect when scrambling code auto detection is OFF for over-the-air
measurements.
The default value for this command is ON. That is, sending SENS:DEM:OTA:LOCK is
equivalent to SENS:DEM:OTA:LOCK ON.
Factory Preset / *RST:
MT8220A: OFF
Front Panel Access:
Measurements, OTA, Code Lock
OTA Sort Parameter
[:SENSe]:OTA:SORT CODE|POWer
[:SENSe]:OTA:SORT?
This command refers to the parameter that is used to sort the over-the-air measurement
results.
Factory Preset / *RST:
MT8220A: CODE
Front Panel Access:
Measurements, OTA, Sort By
PASS/FAIL Test Selection
[:SENSe]:PFail <test set>
[:SENSe]:PFail?
Selects the active test set to be used in subsequent PASS/FAIL measurements. The
<test set> value must correspond to a test set that is defined in the test set list. For
more information on loading test sets to the instrument, refer to the Anritsu Master
Software Tools documentation. NOTE: The first test set in the list would be test set 0, the
next test set would be 1, etc.
When using the query form of the command, the return value will be the currently
selected test set number. If there is not a valid test set selected, the return value will be “-
1” (negative 1).
Front Panel Access:
Measurement, Pass Fail Mode,Select Pass/Fail Test
See Also:
:CONFigure PFail