Default Units: Hertz
See Also:
:CONFigure:DEMod CDP
Measure Spectral Emission Mask
:MEASure:EMISsion?
Sets the active measurement to spectral emission mask, sets the default measurement
parameters, triggers a new measurement and returns the spectral emission mask
measurement results.
It is a combination of the commands:CONFigure:RF
EMISsion;:READ:EMISsion?
For a description of the default spectral emission mask measurement parameters see
:CONFigure:RF EMISsion.
To make a spectral emission mask measurement with settings other than the default
values send
• :CONFigure:RF EMISsion
• Commands to set desired settings
• :READ:EMISsion?
Results are returned as 2 comma-delimited values: pass/fail status, and template number.
See Also:
:MEASure:EMISsion:FAIL?
:MEASure:EMISsion:TEMPlate?
Measure Spectral Emission Mask Pass/Fail status
:MEASure:EMISsion:FAIL?
Sets the active measurement to spectral emission mask, sets the default measurement
parameters, triggers a new measurement and returns the spectral emission mask pass/fail
results.
It is a combination of the commands:CONFigure:RF
SPECMASK;:READ:EMISsion:FAIL?
For a description of the default spectral emission mask measurement parameters see
:CONFigure:RF EMISsion.
To make a spectral emission mask pass/fail status measurement with settings other than
the default values send
• :CONFigure:RF EMISsion
• Commands to set desired settings
• :READ:EMISsion:FAIL?
See Also:
:MEASure:EMISsion
:MEASure:EMISsion:TEMPlate?
:CONFigure:RF EMISsion