Results are returned as 4 comma-delimited values: CPICH power, PCCPCH power,
PICH power, SCCPCH power.
Default Units: dBm
See Also:
:CONFigure:DEMod CDP
Measure Carrier Feed Through
:MEASure:DEMod:CFTHrough?
Sets the active measurement to code domain power, sets the default measurement
parameters, triggers a new measurement and returns the carrier feed through results.
It is a combination of the commands
:CONFigure:DEMod:CDP;:READ:DEMod:CFTHrough?
For a description of the default code domain power measurement parameters see
:CONFigure:DEMod:CDP.
To make a carrier feed through measurement with settings other than the default values
send
• :CONFigure:DEMod CDP
• Commands to set desired settings
• :READ:DEMod:CFTHrough?
Default Units: dB
See Also:
:CONFigure:DEMod CDP
Measure Error Vector Magnitude
:MEASure:DEMod:EVM?
Sets the active measurement to code domain power, sets the default measurement
parameters, triggers a new measurement and returns the EVM results.
It is a combination of the commands
:CONFigure:DEMod:CDP;:READ:DEMod:EVM?
For a description of the default code domain power measurement parameters see
:CONFigure:DEMod:CDP.
To make a carrier feed through measurement with settings other than the default values
send
• :CONFigure:DEMod CDP
• Commands to set desired settings
• :READ:DEMod:EVM?
Default Units: Percent
See Also:
:CONFigure:DEMod CDP