If valid results are not available in the requested position (1-6), the returned string will
contain a series of 3 dashes “---“ instead of the numeric results.
Default Units: Scrambling Code: No units
CPICH power: dBm
Chip Energy: dBm
EC/I0: dB
Pilot dominance: No units
Fetch PASS/FAIL
:FETCh:PFail?
Returns the most recent PASS/FAIL measurement results.
If the instrument is actively measuring, it will not return until the measurement is
complete. If the instrument is not actively measuring and the current data is not valid it
will return error -230. This could occur if there was a *RST immediately before the
:FETCh? or if a measurement parameter was changed without an :INITiate.
Measurement results are returned in a block of ASCII text in the format of
<header><block>. The ASCII header specifies the number of data bytes. It looks
like #AX, where A is the number of digits in X and X is the number of bytes in the
<block>.
The block consists of a set of records which indicate individual test results. Records are
comma-separated. Each record follows the format <test ID>: <input1>
<input2> <min> <max> <value1> <value2> <PASS/FAIL/OPTION
NOT INSTALLED>. The <test ID> field indicates which test was performed to
retrieve these results. The < input1> and < input2> fields indicate the test
setup. The <min> and <max> fields indicate the minimum and maximum values against
which <value1> and <value2> were compared to arrive at the final PASS/FAIL
result.
If the specified test is not available in the instrument, the <PASS/FAIL> field will
indicate “option not available” instead of “pass” or “fail”.
All unused fields for a given test will be represented by a double dash (“--“).
Fetch Peak to Average Power Ratio
:FETCh:PTAPower?
Returns the most recent peak to average power ratio measurement result.
If the instrument is sweeping, it will not return until the sweep is complete. If the
instrument is not sweeping and the current data is not valid it will return error -230. This
could occur if there was a *RST immediately before the :FETCh? or if a measurement
parameter was changed without an :INITiate.
Default Units: dB

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals