Limit Alarm
:CALCulate:LIMit:ALARm OFF|ON|0|1
:CALCulate:LIMit:ALARm?
Enables/disables the currently active limit line alarm.
Setting the value to ON or 1 will turn on the limit alarm.
Setting the value to OFF or 0 will turn off the limit alarm.
The query version of the command returns a 1 if the currently selected limit line alarm is
set to ON and returns 0 if OFF.
Use :CALCulate:LIMit:TYPe to set the currently active limit line.
Factory Preset / *RST:
OFF
Front Panel Access:
Shift-6 (Limit), Limit Alarm
See Also:
:CALCulate:LIMit:TYPe
Lower Limit Alarm
:CALCulate:LIMit:LOWer:ALARm OFF|ON|0|1
:CALCulate:LIMit:LOWer:ALARm?
Enables/disables the lower limit alarm.
It is a combination of the commands :CALCulate:LIMit:TYPe 1;
:CALCulate:LIMit:ALARm ON|OFF.
Setting the value to ON or 1 will turn on the lower limit alarm.
Setting the value to OFF or 0 will turn off the lower limit alarm.
The query version of the command returns a 1 if the lower limit line alarm is ON and
returns 0 if OFF.
Note that using this command set the lower limit line active for editing.
Factory Preset / *RST:
OFF
Front Panel Access:
Shift-6 (Limit), Limit Alarm
See Also:
:CALCulate:LIMit:ALARm
Number of Lower Limit Points
:CALCulate:LIMit:LOWer:POINt?
Returns the number of points currently in the lower limit line.
Note that issuing this command will set the lower limit line active for editing
(:CALCulate:LIMit:TYPe 1).
Lower Limit State
:CALCulate:LIMit:LOWer[:STATe] OFF|ON|0|1
:CALCulate:LIMit:LOWer[:STATe]?
Turns the lower limit line ON or OFF.