Reduction ratio selection.
or
WARNING!
An icon appears next to the torque value identifying the reading
tolerance on the value indicated:
tolerance of ±10%
tolerance of ±20%.
NOTE: the data is automatically stored.
Removable cord
Cleaning and care.
It is recommended to use Daily Oil (CEFLA s.c.) for lubrication.
WARNING!
• Do not soak the micromotor in liquid disinfectants or detergents.
• Products not recommended: harsh products and/or products containing acetone, chlorine and sodium hypochlorites.
Sterilization.
WARNING!
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
Instructions for use.
WARNING!
• The instrument is supplied NOT STERILE and must be sterilized before use (see paragraph 1.5.).
Before performing sterilization, check the instructions for use included with the device.
• Only for American and Canadian markets: instruments must be FDA-approved.
• Never put the contra angle on the micromotor while it is running.
• The chuck release button must be held down during operation!
Friction between the button and micromotor rotor overheats the head and may cause burns.
• The patient’s internal tissues (tongue, cheeks, lips, etc...) must be protected against contact with the button by using suitable instruments
(mirror, etc...).
• The chucks and various instruments attached to the handpieces must comply to the standard ISO 10993.