66
66
EN
R7
- OPERATING INSTRUCTION
7.3. Manual SHS system
Description of the system
board.
WARNING!
With the manual SHS system, the tank reserve ( a ) and SHS system
active/inactive warning icons are not shown on the TOUCH DISPLAY.
WARNING!
Perform a disinfecting cycle at the end of each work day.
Filling the tank.
• c
a
NOTE: while this operation is being performed, the pressurised
air in the tank is automatically released outside.
WARNING!
Use only distilled water. For a higher guarantee of hygiene you can
add 600 parts per million (ppm) of hydrogen peroxide using 20 ml of
Peroxy Ag+ (not available for the American and Canadian markets) per
litre of distilled water, or oxygenated water (20 ml of 3% oxygenated
water per litre of distilled water).
c
WARNING!
period of time (holiday), always empty the hydrogen peroxide tank
( a ) before leaving.
BY-PASS function
b
Cleaning the tank.